33II EN
1. Set the function switch at desired current position.
2. Open Spring-loaded clamp by pressing trigger on right
side of meter.
3. Position clamp around wire or conductor and release
clamp trigger, make sure that the clamp is entirely closed.
The clamp must be positioned around only one conductors
of a circuit . If the clamp is placed around two or more
current carrying conductors, the meter reads FALSE.
1. Set the function switch at " V ~ " position.
2. Connect the black test lead to the "COM" terminal and the
red test lead to the " V-Ω" terminal.
You can now place the test probes on the conductors to
make the measurement.
1. Set the function switch at " Ω " position.
2. Connect the black test lead to the "COM" terminal and red
test lead to the " V– Ω " terminal.
3. Verify that the power to the circuit under test is o.
Connect test leads to the circuit to make the measurement.
4. Built in buzzer sounds if the resistance of the circuit under
test is less than approximately 50Ω.
7. If the meter is used near equipment that generates
electromagnetic noise, the display may be unstable or
indicate large errors.
8. Maximum rated voltage to earth for voltage
measurements terminals is 600V CATIII.
9. When using the instrument as a Voltmeter or ohmmeter
never clamp the jaws around or onto a conductor.
THIS INSTRUMENT MUST NOT BE USED ON
UNINSULATED CONDUCTORS AT A VOLTAGE
GREATER THAN 600V ac/dc.
3-2 AC Current Measurement
3-3 AC Voltage Measurement
3-4 Resistance Measurement