Compex FIT 1.0, FIT 3.0, SP 2.0 & SP 4.0 Руководство пользователя

Тип
Руководство пользователя
Instruction for use
Mode d’emploi
Anweisungen
Istruzioni
Instrucciones
Instructies
Instruções
Инструкция
Bruksanvisning
Instruction for use
01
EN
TABLE OF CONTENTS
1. Explanation of symbols 02
2. How does electrostimulation work? 03
3. How does MI (Muscle Intelligence) technology work? 05
4. Device function 06
Contents of kits and accessories 06
Device description 07
Battery insertion 08
Connection 08
Preliminary settings 09
Choosing a category 09
Selecting a programme 10
MI-scan test 11
Adjusting stimulation intensities 11
Programme progression 12
End of a programme 13
Battery level and charging 13
5. Troubleshooting 15
6. Device maintenance 18
7. Technical specifications 19
8. EMC Table 21
It is strongly recommended to read these instructions and the contra-indications and safety
measures carefully before using your stimulator.
02
EN
1. EXPLANATION OF SYMBOLS
See the instructions
The stimulator is a category II device with built-in power supply and type BF applied parts.
Manufacturer’s name and address and date of manufacture
Name and address of approved representative in the European Union
This device must be separated from household waste and sent to special collection facilities for
recycling and recovery
The stand-by button is multi-functional
Protect from sunlight
Store in a dry place
This is an indication for protection against ingress of water and particulate matter. The mark IP20
on your unit means: your unit is protected against solid foreign objects of 12.5mm dia and greater.
Not protected against water
IP02 on the carrying case means: Protected from the ingress of water droplets from a shower
of rain.
Latex-free
Reference number
Batch number
IP02
on the case
IP20
on the unit
03
EN
2. HOW DOES ELECTROSTIMULATION WORK?
Electrostimulation involves stimulating nerve fibres by electrical impulses transmitted by electrodes. The
electrical impulses produced by Compex stimulators are high-quality impulses that are safe, comfortable
and effective and stimulate various types of nerve fibres:
1. Motor nerves to stimulate a muscular response, referred to as electrical muscle stimulation (EMS).
2. Certain types of sensitive nerve fibres to obtain analgesic effects or pain relief.
1. STIMULATION OF MOTOR NERVES (EMS)
With voluntary activity, the brain orders muscles to contract and a command is then sent to nerve fibres
in the form of an electrical signal. This signal is then sent to muscle fibres, which contract. The principle of
electrostimulation accurately reproduces the process involved in a voluntary contraction. The stimulator
sends an electrical impulse to nerve fibres to excite them. This excitation is then transmitted to muscle
fibres and results in a basic mechanical response (= a twitch). This is the basic requirement for muscular
contraction. The muscular response is to all intents and purposes identical to the muscular work
controlled by the brain. In other words, the muscle does not distinguish between a command sent by the
brain or the stimulator.
Programme settings (number of impulses per second, duration of contraction, rest time, total programme
duration) subject the muscle to various types of work, depending on the muscle fibre. Various types
of muscle fibres can be identified according to their respective contraction speeds: slow, intermediate
and fast fibres. A sprinter clearly has more fast fibres and a marathon runner has more slow fibres. With
good knowledge of human physiology and full control of stimulation settings of the various programmes,
muscular work can be specifically pinpointed to achieve the desired objective (muscular strengthening,
increased blood circulation, firming, etc.).
04
EN
2. STIMULATION OF SENSORY NERVES
Electrical impulses can also excite sensory nerve fibres to obtain analgesic effects or pain relief.
Stimulating tactile sensory nerve fibres blocks pain being transmitted to the nervous system. Stimulating
another type of sensory fibre increases the production of endorphins and therefore reduces pain. With
pain relief programmes, electrostimulation can be used to treat acute or chronic localised pain and
muscle pain.
Caution: Do not use pain relief programmes for an extended period without medical advice.
BENEFITS OF ELECTROSTIMULATION
Electrostimulation is a very effective method for making muscles work:
• With a significant improvement in various muscular qualities
• With no cardiovascular or mental fatigue
• With limited stress exerted on joints and tendons. Electrostimulation thereby enables more muscular
work than voluntary activity.
For optimum results, Compex recommends supplementing your electrostimulation sessions with other
commitments, such as:
• Regular physical exercise
• A balanced and healthy diet
• A balanced lifestyle
05
EN
3. HOW DOES MI (MUSCLE INTELLIGENCE)
TECHNOLOGY WORK?
To access MI functions the MI-sensor cable (not available on all devices) must be connected to the stimulator.
MI-SCAN
Just before starting a work session, the MI-scan function probes the chosen muscle group and automatically
adjusts the stimulator settings to the excitability of this area of the body, depending on your physiology.
This function results in a short test sequence at the start of the programme, during which measurements
are taken.
At the end of the test, the intensity has to be increased to start the programme.
MI-TENS
The MI-tens function limits unwanted muscle contractions in painful areas.
With each intensity increase applied by the user, a test phase occurs, and if a muscular contraction is
detected, the device automatically reduces the intensity of the stimulation.
This function is only accessible in the TENS, Epicondylitis and Tendinitis programmes.
MI-RANGE
The MI-range function indicates the ideal range for adjusting stimulation intensity, to be followed in the
following programmes: recovery, massage, capillarisation or even muscle pain.
When the device has detected the ideal intensity range, check mark will appear on the screen. For optimum
work, intensity should be kept within this range.
Not all devices have MI technology. The table below shows which functions are available with each device.
SP 2.0 SP 4.0 FIT 1.0 FIT 3.0
MI-SCAN
-
MI-TENS -
-
-
MI-RANGE -
-
-
06
EN
4. DEVICE FUNCTION
CONTENTS OF KITS AND ACCESSORIES
SP 2.0 SP 4.0 FIT 1.0 FIT 3.0
REF QTY REF QTY REF QTY REF QTY
STIMULATOR 001096 1 001095 1 001098 1 001097 1
CHARGER 6830XX 1 6830XX 1 6830XX 1 6830XX 1
SET OF 4 SNAP CABLES 001119 1 001119 1 001119 1 001119 1
MI-SENSOR CABLE 601160 1 601160 1 N/A N/A 601160 1
BAG OF SMALL 5X5
ELECTRODES
42215 2 42215 2 42215 2 42215 2
BAG OF LARGE 5X10
ELECTRODES
42216 2 42216 2 42216 2 42216 2
INSTRUCTIONS ON
CD-ROM
880053 1 880053 1 880053 1 880053 1
QUICK START GUIDE 885624 1 885624 1 885624 1 885624 1
CARRY CASE 680029 1 680029 1 680029 1 680029 1
BATTERY PACK 94121X 1 94121X 1 94121X 1 94121X 1
Only use this device with cables, electrodes, battery, power adaptor and accessories
recommended by Compex.
07
EN
DEVICE DESCRIPTION
A Standby button
B Sockets for the 4 stimulation cables
C Stimulation cables
D +/- buttons for the 4 stimulation channels
E I-button, which allows:
• The intensities to be increased on several channels at the same time
• The last 5 programmes used to be accessed
F Charger socket (slide the red cover to the right to reveal the charger connector)
G Battery compartment
H Socket for belt clip
A
B
C
DE
F
G
H
08
EN
BATTERY INSERTION
Open the battery compartment cover and insert the battery, with the label pointing upwards so that the +
and - terminals are opposite the device’s contacts. Then replace the cover. If the device is not going to be
used for over 3 months, ensure that the battery is fully charged. If the device is not going to be used for over
6 months, ensure that the battery is fully charged and remove it from the stimulator. Switch the stimulator off
before removing the battery.
CONNECTIONS
CHARGER CONNECTION
Remove all stimulation cables from the stimulator before recharging it. Connect the charger to a wall socket
and connect the stimulator by sliding the red cover to the right to reveal the charger connector. It is strongly
recommended to fully charge the battery before its first use to improve its performance and life span.
CONNECTING THE CABLES
The cables of the stimulator connect to the 4 sockets at the front of the device. The MI-sensor cable (if
available with the device) can be connected to any socket on the stimulator.
COMPEX
COMPEX
09
EN
PRELIMINARY SETTINGS
Before your first use, you can define various settings. This options screen can then be displayed by switching
the device off and by pressing and holding the on/off button for at least 2 seconds.
B Press the +/- button of channel 1 to select the language to be used.
C Press the +/- button of channel 2 to adjust screen contrast.
D Use the +/- button of channel 3 to adjust the volume.
E Press the +/- button of channel 4 to adjust the backlighting.
ON: backlighting always active.
OFF: backlighting always inactive.
AUTO: backlighting activated every time a button is pressed.
A Press the on/off button to confirm and save your selections. Settings are immediately applied.
CHOOSING A CATEGORY
N.B. The following screens are generic examples but they work in the same way regardless of the device that
you have.
To switch the stimulator on, briefly press the on/off button.
Before selecting a programme, you should select the desired category.
A Press the on/off button to switch the device off.
B Press the +/- button of channel 1 to select a category.
E Press the +/- button of channel 4 to confirm your selection.
A B C D E
A B E
10
EN
TOP 5
To access the last 5 programmes used press the I-button. From there, select the desired programme and
start it.
SELECTING A PROGRAMME
A Press the on/off button to return to the previous screen.
B Press the +/- button of channel 1 to select a programme.
E Press the +/- button of channel 4 to confirm the selection and start the stimulation session.
N.B. After selecting a programme an additional screen may appear, requesting the programme cycle to be
selected. Select the desired cycle and then press the +/- button of channel 4 to confirm the selection and
start the stimulation session.
It is recommended to start with the 1st cycle and to change cycle when it is complete, normally after 4
to 6 weeks of stimulation based on 3 sessions per week. It is also important to have reached significant
stimulation intensities during the sessions before progressing to another cycle.
At the end of the cycle you can either start a new cycle or carry out maintenance training based on one
session per week.
A B E
11
EN
MI-SCAN TEST
N.B. See the chapter “How MI technology works”.
If the MI-sensor cable is connected, the MI-scan test starts immediately after the programme is selected.
A Press the on/off button to stop the test
The +/- buttons of the 4 channels are inactive throughout the test.
ADJUSTING STIMULATION INTENSITIES
When you start a programme, you are asked to increase the stimulation intensities.
This step is essential for a successful session.
1 Programme duration in minutes and seconds
2 Programme progression bar. For details of how it works, see the following paragraph:
“Programme progression”
A Press the on/off button to switch the unit into Pause mode.
B C D E The four channels flash, going from + to 000. The stimulation intensity must be increased to be able
to start stimulation. For this, press the + buttons of the relevant channels until the desired setting is reached.
N.B. To increase the intensities on several channels simultaneously, press the I-button and then increase the
intensities. The interdependent channels are displayed in white on a black background.
A
A B C D E
1
2
12
EN
A B C D E
PROGRAMME PROGRESSION
Stimulation starts properly once the stimulation intensity has been increased.
The examples below explain the general rules. Depending on the programme, there may be slight differences.
1 Time remaining (in minutes and seconds) until the end of a programme
2 The duration bar showing the duration of the contraction and duration of active rest is only shown during
the work sequence
3 Session sequences
4 Warm-up
5 Work period
6 Relaxation
A Press the on/off button to temporarily interrupt the programme. To resume, simply press the +/- button of
channel 4. The session will resume at 80% of the intensity level being used before it was interrupted.
N.B. In pause mode, the button switches directly to the next sequence.
N.B. In pause mode and depending on the programme, usage statistics may be displayed:
MAX = the maximum intensity reached per channel during contraction phases
AVG= the average intensity for all channels applied during the contraction phases
B C D E The various intensities reached during the contraction phase are shown by a series of black vertical
bars; the intensities of the rest phase are illustrated by hatched bars.
Please note that the stimulation intensities for the active rest phase are automatically set to 50% of the
contraction intensities. They can be changed during the rest phase. Once changed, they are completely
independent of the contraction intensities.
1
23
4 5 6
13
EN
END OF A PROGRAMME
At the end of a session, the following screen is displayed. To stop the stimulator, press the on/off button.
N.B. Depending on the programme, usage statistics may be displayed (see previous chapter
“Programme progression”).
BATTERY LEVEL AND CHARGING
Battery performance depends on the programme and the stimulation intensity applied. It is strongly
recommended to fully charge the battery before its first use to improve its performance and life span. Always
use the charger supplied by Compex to recharge the battery.
If the device is not going to be used for over 3 months, ensure that the battery is fully charged. If the device
is not going to be used for over 6 months, ensure that the battery is fully charged and remove it from the
stimulator. Switch the stimulator off before removing the battery.
BATTERY LEVEL
The battery charge level is shown by a battery icon in the bottom left of the screen. The battery icon flashes
when the battery is completely flat. The device can then no longer be used. Recharge it immediately.
RECHARGING
Remove all stimulation cables from the stimulator before recharging it. Connect the charger to a wall socket
and connect the stimulator by sliding the red cover to the right to reveal the charger connector.
14
EN
The charging menu shown below appears automatically.
The charging duration is displayed on the screen. As soon as charging is complete, the battery flashes.
Disconnect the charger: the stimulator will switch off automatically.
15
EN
5. TROUBLESHOOTING
ELECTRODE FAULT
The device emits a tone and alternately displays the pair of electrodes symbol and an arrow pointing towards
the channel on which a problem has been detected. In the above example, the stimulator has detected an
error on channel 1.
Check that the electrodes are connected to this channel.
If electrodes are old, worn and/or if the contact is poor: try using new electrodes.
Try using the stimulation cable on a different channel. If the cable is still faulty, replace it
(www.compexstore.com).
STIMULATION DOES NOT PRODUCE THE USUAL SENSATION
Check that all settings are correct and check electrodes are properly positioned.
Change the position of the electrodes slightly.
STIMULATION CAUSES DISCOMFORT
Electrodes lose their adhesive capacity and no longer provide suitable contact with skin.
Electrodes are worn and must be replaced.
Change the position of the electrodes slightly.
16
EN
THE STIMULATOR IS NOT WORKING
When an error screen is displayed during use, note the error number (in the example, the error number is
1/0/0) and contact the Compex-approved customer service centre.
THE BATTERY IS VERY FLAT
If the following screen appears, switch the device off and connect the charger.
If a battery is very flat, a recovery cycle will start, lasting 2 minutes.
17
EN
When this has finished, if the battery is working properly, it will start charging; in this case, it is strongly
recommended to perform a charge/discharge cycle for the battery by pressing on the channel 4 button
to launch this cycle, which may take up to 12 hours. If, however, it proves to be faulty, the following screen
appears and the battery must be replaced.
18
EN
6. DEVICE MAINTENANCE
GUARANTEE
See the attached leaflet.
MAINTENANCE
Your stimulator does not require neither calibration nor periodic maintenance. However when needed clean
with a soft cloth and solvent-free alcohol-based detergent. Use as little liquid as possible to clean the device.
Do not disassemble the stimulator or the charger as they contain high-voltage components which could
cause electrocution. This must be carried out by Compex-approved technicians or repair services. If your
stimulator contains parts that appear to be worn or faulty, please contact the closest Compex customer
service centre.
CONDITIONS FOR STORAGE/TRANSPORT AND USE
STORAGE AND TRANSPORT USE
TEMPERATURE -20° C to 45° C 0° C to 40° C
MAXIMUM RELATIVE HUMIDITY 75% 30% to 75%
ATMOSPHERIC PRESSURE from 700 hPa to 1060 hPa from 700 hPa to 1060 hPa
Do not use in areas at risk of explosion.
DISPOSAL
Batteries must be disposed of in accordance with national regulatory requirements in force. Any product
bearing the WEEE label (a crossed-out wheeled bin) must be separated from household waste and sent to
special collection facilities for recycling and recovery.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226

Compex FIT 1.0, FIT 3.0, SP 2.0 & SP 4.0 Руководство пользователя

Тип
Руководство пользователя

Задайте вопрос, и я найду ответ в документе

Поиск информации в документе стал проще с помощью ИИ